Designed for a wide range
of pre-hospital, hospital or
primary care needs, the Welch Allyn PIC 40 provides
everything needed for patient monitoring and defibrillating.

The Welch Allyn PIC 40 is a combination vital signs monitor and defibrillator
ready to deliver immediate performance wherever sudden cardiac arrest may
strike.
The PIC 40 comes equipped with defibrillation capabilities, ECG monitoring,
voice-memo, synchronized cardioversion, Nellcor SpO2, pacing, optional NIBP and
optional data record and playback.
This breadth of options makes the Welch Allyn PIC 40 the first choice for
multiple levels of emergency patient care. Ready for rescue, the versatile Welch
Allyn PIC 40 provides dependable emergency/monitoring performance for both the
first responder and advanced team. The instrument features an integrated
Automated External Defibrillator (AED) for use by immediate care responders.
The PIC 40 is also upgradeable. Increasing your treatment potential and purchase
value without increasing the physical size and weight of the unit.
• Biphasic
• Manual defibrillation (2 - 360 joules)
• Synchronized cardioversion
• Automated External Defibrillation with escalating energy of 200/300/360 (voice
and text prompts)
• Nellcor SpO2
• External Pacing
• 6.5" color display
• High Performance Welch Allyn SuperPac™ rechargeable batteries using
sophisticated NiMH technology
• Innovative design to allow for future upgrades
• Simple operation with dedicated treatment parameter controls
Optional Features
• NIBP
• Data record and playback
GENERAL
Weight: 10.5 lbs (4.76 kg) basic configuration.
Size: 16.7 x 12.5 x 5.3 inches; 41.75 x 32 x 13 cm.
Operating:
Time: 2 hours
Temperature*: 0 to 45° C
Humidity* (NC): 15 to 95% RH
Vibration*: MIL-STD 810E
Shock / Drop*: MIL-STD 810E
Altitude: Up to 15,000 feet
Transport and Storage: Temperature (-30 to 70˚C),
humidity (15 to 95% non-condensing), atmospheric
pressure ( 860 to 1060 hPa), shock/vibration (ISTA 1A)
Enclosure Protection*:
Solid Foreign Object: IEC 529, IP2X
Water: IEC 529, IPX4.
MONITOR/DISPLAY
Input: 3-lead or 5-lead patient cable, paddles, or
multipurpose hands-free adapter.
Size: 6.4 inch (16.5 cm) diagonal, non-fade.
Type: LCD color active matrix.
Display Resolution: 640 x 480 pixels.
Sweep Speed: 25 mm / sec.
Lead Selections: Paddles (Pads), I, II, III, AVR, AVL,
AVF, V.
Frequency Response: (User-selectable).
• 2 to 20 Hz Limited mode
• 0.5 to 40 Hz Monitor mode
• 0.05 to 150 Hz Diagnostic mode(automatically sets chart recorder response)
Common Mode Rejection: Complies with AAMI EC13-
1992 section 3.2.9.10.
Heart Rate Meter: 20 to 300 BPM.
Heart Rate Alarms: User-selectable.
Size: 0.125, 0.25, 0.5, 1, 2, 4 cm/mv and auto-ranging.
AED
Configurable: Basic & Basic+ modes
Operation: Continuous analysis of ECG advises shock
on VFIB, shockable VTACH
Energy Sequence: User configurable
Biphasic: 150/200/300/360J
Factory Default:
Biphasic: 150/200/300J
Prompts: Audible voice and visual text prompts
DEFIBRILLATOR
Waveform:
Biphasic: Low current, impedance compensated,
optimized terminating current, truncated exponential
biphasic
Energy Select:
External: 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 J.
Internal: 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50 J.
Charge Time: < 10 seconds @ 360 J (typical) SuperPak
Output: Adult paddles, internal paddles, peds adapter
and multipurpose hands-free adapter available.
Synchronizer: Delivers energy within 60 msec.
Disarm: Front panel switch.
RECORDER
Type: High-resolution thermal array.
Annotation: Time, date, ECG lead, ECG gain, heart rate,
defibrillation and pacing parameters and treatment
summary ACLS events.
Paper Width: 50 mm.
Paper Speed: 25 mm/sec.
Delay: 6 seconds.
Frequency Response: Automatically set to monitor's
frequency response.
Tx Summary Log: 60 ECG events or 300 non-ECG events.
BATTERY
High Capacity Type: NiMH 12 volt - Welch Allyn
SuperPac™.
Low Battery Indicator: Flashing low battery icon on
display.
Multiple Paired Contacts: Ensures quick, error-free
insertion and backup reliability.
Recharge Rate: 80% in 7.5 hours, 100% in 9.5 hours.
(For high capacity batteries)
PULSE OXIMETER
Measurement Range: 31 to 99%.
Heart Rate Range: 30 to 250 BPM ±5% or 5 BPM whichever is less.
Accuracy: 81-99% = ±2%, 70-80% = 3%, 0-69% unspecified
Update Period: Every pulse after 4 valid pulses
Alarm Limits: 60 lower limit, 70 upper limit to 99%
Probe: Reusable finger or multi-site probe |